CSA C22.2 NO. 80601-2-60:14 (R2019):2014 Edition
$50.05
Medical electrical equipment – Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
Published By | Publication Date | Number of Pages |
CSA | 2014-10-01 | 94 |
CSA Preface
This is the first edition of CAN/CSA-C22.2 No. 80601-2-60, Medical electrical equipment – Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment, which is an adoption, with Canadian deviations, of the identically titled IEC (International Electrotechnical Commission) Standard 80601-2-60 (first edition, 2012-02). It is one in a series of Standards issued by CSA Group under Part II of the Canadian Electrical Code.
This Standard is intended to be used in conjunction with CAN/CSA-C22.2 No. 60601-1:14, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance (adopted IEC 60601-1:2005+A1:2012, with Canadian deviations).
This Standard has been developed in compliance with Standards Council of Canada requirements for National Standards of Canada. It has been published as a National Standard of Canada by CSA Group.
201.1 Scope, object and related standards
Clause 1 of the general standard1) applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of DENTAL UNITS, DENTAL PATIENT CHAIRS, DENTAL HANDPIECES and DENTAL OPERATING LIGHTS, hereafter referred to as DENTAL EQUIPMENT.
Excluded are amalgamators, sterilizers and dental X-ray equipment.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
201.1.2 Object
Replacement:
The object of this particular standard is to establish BASIC SAFETY and ESSENTIAL PERFORMANCE requirements for DENTAL EQUIPMENT (as defined in 201.3.202.)